User and Programming Manual
90
XDS
Series
11.3 Source Subsystem
This subsystem programs all the output parameters of the power supply.
Subsystem Syntax
[SOURce:]
CURRent:
[LEVel]
Set
the
current
value
in
amps.
HIGH
Set
the
maximum
current
allowed
LOW
Set
the
minimum
current
allowed
LIMit:
CURRent?
Current
range
limt:
Model
specific
VOLTage?
Voltage
range
limit.
Model
specific
MODE?
Returns
power
supply
operating
mode.
VOLTage:
[LEVel]
Set
the
output
voltage
value
in
volts.
HIGH
Set
the
maximum
voltage
allowed
LOW
Set
the
minimum
voltage
allowed
PROTection
[:LEVel]
Set the over voltage protection trip level in % of
range.
STATe
Enables or disabels OVP mode.
[SOURce:]CURRent[:LEVel]
This command sets the current limit of the output. If the load current exceeds the programmed
current limit, the power supply will cross over from voltage mode to current mode. In this mode, the
output voltage will fold back as much as needed to retain the current at the programmed limit.
The current may be set between the low limit and high limit as programmed using the CURR:HIGH
and CURR:LOW commands. However, the high limit cannot be higher than the model's capability
(configuration limit) which can be queried with the LIM:CURR? command. The low limit cannot be
less than zero. The *RST state for high and low current is zero for low, and maximum configuration
limit for high. (model specific).
Command Syntax
[SOURce:]CURRent <NRf>
Parameters
value between CURR:HIGH and CURR:LOW settings.
Unit A
(amperes)
*RST
Defined by the PON
Examples CURR
5
Query Syntax
CURRent?
Returned Parameters
<NR2>
Related Commands
CURR:HIGH
CURR:LOW
[SOURce:]CURRent:HIGH
This command sets the maximum programmed current that will be accepted by the power supply. If a
value above this limit value is programmed, a "-200 Execution Error" will be generated.
The current limit may be set between zero and the model's capability (configuration limit) which can
be queried with the LIM:CURR? command. The high limit cannot be less than the low limit however.
The *RST state for high current limit is the maximum configuration limit. (model specific).
Summary of Contents for XDS 100-100
Page 4: ...User and Programming Manual 4 XDS Series Electrical Safety Symbols Used in This Manual ...
Page 27: ...User and Programming Manual XDS Series 27 Figure 3 5 Outline Drawing XDS DC Supply ...
Page 41: ...User and Programming Manual XDS Series 41 Figure 5 4 Parallel Cable Interconnect Accessory ...
Page 69: ...User and Programming Manual XDS Series 69 Figure 8 3 Replaceable Fuse Locations ...