User and Programming Manual
92
XDS
Series
[SOURce:]MODE
This command is used to query the power supply's operating mode. The power supply can operate
in either Constant Voltage or Constant Current mode. A numeric value is returned for a MODE?
Query and is encoded as shown in the table below:
Value
Mode
0 Undefined.
1
Constant Voltage (CV)
2
Constant Current (CC)
3
Cross over CV and CC mode
Table 11-1: Mode query result decoding
Command Syntax
[SOURce:]MODE?
Parameters None
Query Syntax
MODE?
Returned Parameters
<NR1>
Related Commands
none
[SOURce:]VOLTage[:LEVel]
This command programs the DC output voltage level of the power supply.
The voltage may be set between the low limit and high limit as programmed using the VOLT:HIGH
and VOLT:LOW commands. However, the high limit cannot be higher than the model's capability
(configuration limit) which can be queried with the LIM:VOLT? command. The low limit cannot be
less than zero. The *RST state for high and low voltage is zero for low, and maximum configuration
limit voltage for high. (model specific).
Command Syntax
[SOURce:]VOLTage[:LEVel] <NRf>
Parameters
value between VOLT:HIGH and VOLT:LOW settings.
Unit
V (rms voltage)
*RST Value
0 volt
Examples
VOLT 250
VOLT:LEV 25
Query Syntax
[SOURce:]VOLTage[:LEVel]?
Returned Parameters
<NR2>
Related Commands
VOLT:HIGH
VOLT:LOW
Summary of Contents for XDS 100-100
Page 4: ...User and Programming Manual 4 XDS Series Electrical Safety Symbols Used in This Manual ...
Page 27: ...User and Programming Manual XDS Series 27 Figure 3 5 Outline Drawing XDS DC Supply ...
Page 41: ...User and Programming Manual XDS Series 41 Figure 5 4 Parallel Cable Interconnect Accessory ...
Page 69: ...User and Programming Manual XDS Series 69 Figure 8 3 Replaceable Fuse Locations ...