9-5
9-3 Initialization
You will lose all the setting data stored in the memory by initializing. The lost data can not be
retrieved.
Initialization swaps all the data stored in the memory in the RAM and the EEPROM with the default
data that factory has configured originally. There are 3 types of initialization depending on items to be
initialized. The item marked with “
√
” in the table below will be initialized.
Initializing Item
Initialization Mode
RAM Stores;
EEPROM Stores;
Zero
Tare
Set point
Total weight
Total count
Function
data
Calibration
data
Mode name
Display in the
initialization mode
√
RAM initialization
“init rA”
√
√
RAM & Function initialization
“initFnc”
√
√
√
All data initialization
“initAll”
Initialization procedures
First make sure that the calibration disable switch (inside the case) is set to the ON position. If not,
initialization will not be performed.
Select the type of initialization and follow the procedure number to initialize.
*
To return to the normal mode in the middle of initializing, press [ESC].
RAM initialization:
“init rA”
RAM & Function initialization:
“init Fnc”
All data initialization:
“initALL”
Enter the RAM initialization mode
1
Turn the power on while pressing [SET POINT].
“init rA”
Is displayed.
Enter the RAM & Function initialization mode
1
Turn the power on while pressing [SET POINT].
“init rA”
Is displayed.
Press [F],
“init Fnc”
is displayed.
Enter the All data initialization mode
1
Press [CALF] while pressing [ENTER] in the normal mode.
Press [ENTER].
Press [F],
“initALL”
is displayed.
Initialization
2 2 2
Press [ENTER], display blinks to warn against losing data.
If not initialize
Press [ESC] to escape from the initialization mode without performing and return to
the normal mode.
To initialize
Press [ENTER] for 3 seconds to initialize the data. It will initialize the data and returns
to the normal mode.
Summary of Contents for Explosion Protected Weighing Indicator AD-4403-FP ST Series
Page 22: ...2 7 Blank page...
Page 36: ...6 3 Blank page...
Page 38: ...6 5...
Page 40: ...6 7...
Page 53: ...6 20 Blank page...
Page 73: ...8 18 Blank page...
Page 83: ...Dimensions OP 20 Stand A 5...
Page 85: ...A 7 Blank page...
Page 87: ...B 2 Blank page...
Page 105: ...Fig Apx D 02 Gravity map WMF D 2...
Page 107: ...Form F Factory setting rSF 02 5 Command mode Address None rSF 09 0 Address 1 rSF 09 1 E 2...