3-2
3-3 Setting and Recalling Set points
Setting set points
1. Enter the set point setting mode.
Press [CODE] [
code number (2 digits)
] [SET POINT] [ENTER] in this order. It shows;
Main
display:
“CodE XX”
Sub
display:
“YYYYYY”
Status display section (lower): “FINAL” mark turns on.
Left LEDs: “SET POINT” LED blinks.
*
Use numerical keys ([0] – [9]) to type a code number.
*
Set a blank for the code number with [+/-] to change the set point in use.
*
Press [ESC] to return to the normal mode.
2. Select a set point.
Select a set point with [F].
3. Input a set point.
Input a set point with numerical keys ([0] – [9]) and set polarity with [+/-].
*
Preset tare weight can be set up to the capacity. If exceeds,
“dAtA Err”
appears on the sub
display for 3 seconds. Try it again.
*
Press [ESC] to cancel the input. It returns to the previous value. (While the previous value is
displayed, press [ESC] to return to the normal mode.)
*
[+/-] may not work with some set points.
4. Enter the set point into memory.
Press [ENTER] to store the set point into memory.
*
Settings won’t complete until [ENTER] is pressed.
Recalling set points
1. Input a code number to recall.
Press [CODE] [
code number desired (2 digits)
] [ENTER] in this order.
*
Use numerical keys ([0] – [9]) to type a code number.
*
Press [ESC] to cancel the input and re-input. Or erase the digits typed with [+/-] and continue
typing.
*
Code numbers without the final weight setting can’t be recalled. In this case,
“FinAL ”
and
“not dAtA”
are displayed for 3 seconds and returns to the normal mode.
Summary of Contents for Explosion Protected Weighing Indicator AD-4403-FP ST Series
Page 22: ...2 7 Blank page...
Page 36: ...6 3 Blank page...
Page 38: ...6 5...
Page 40: ...6 7...
Page 53: ...6 20 Blank page...
Page 73: ...8 18 Blank page...
Page 83: ...Dimensions OP 20 Stand A 5...
Page 85: ...A 7 Blank page...
Page 87: ...B 2 Blank page...
Page 105: ...Fig Apx D 02 Gravity map WMF D 2...
Page 107: ...Form F Factory setting rSF 02 5 Command mode Address None rSF 09 0 Address 1 rSF 09 1 E 2...