![AND Explosion Protected Weighing Indicator AD-4403-FP ST Series Instruction Manual Download Page 70](http://html1.mh-extra.com/html/and/explosion-protected-weighing-indicator-ad-4403-fp-st-series/explosion-protected-weighing-indicator-ad-4403-fp-st-series_instruction-manual_2941315070.webp)
8-15
*
See {8-6 Data Transmitting Format —Data Format} for “Data Format”.
*
See {8-7 Command Mode— Communication procedure} for “Procedure”.
Command Command name
Function
Procedure
Data
Format
HB
(Halt Batch)
Halt batch in an emergency
A
•
This command is available only in the built-in automatic program mode weighing.
•
In other weighing mode, "IE" will be transmitted.
RF
(Request Final)
Recalls the final weight
B
A/B
Outputs the net weight at batch finish.
RB
(Req.Batch
status)
Recalls the weighing sequence
status
B
H
RTXX
(Request Total)
Recalls the accumulated data of
the code designated by XX
B
C/D
•
When no code designated (blank), total accumulated weight and the total accumulated count will be
recalled.
DTXX
(Delete Total)
Clears the accumulation data of
the code designated by XX
A
When no code designated (blank), total accumulated weight and the total accumulated count will be
deleted.
ET
(Erase Total)
Clears the accumulation data of
all the codes
A
Both total accumulated weight and total accumulated count are also cleared.
RE
(Read EEPROM)
Recalls the data from the
EEPROM
B
•
The EEPROM stores data related to calibration and functions.
•
The data read out with this command can be exported to other Indicator with the ”WE” command
•
This command is available only when
“CAL SEt”
is displayed in the calibration mode.
•
Data length: 308 characters, ASCII code (terminators not included.)
WE
(Write EEPROM)
Writes data (calibration, function)
into the EEPROM
C
This command is available only when
“CAL SEt”
is displayed in the calibration mode.
RC
(Read Code)
Recalls the data of all the codes
B
G
•
Set point and accumulation data of the code are read out from the RAM.
•
The data of the code number from “0” to “99” and the set point data now in use are output in this
order.
•
This command is available only when
“opErAte”
is displayed in the set point editing mode.
WC XX
(Write Code)
Write data to the code number
designated by XX
C
G
This command is available only
“opErAte”
is displayed in the set point editing mode.
RD
(Read function
Data)
Recalls all the function data
D
F
This command is available only when
“opErAte”
is displayed in the set point editing mode.
WD
(Write function
Data)
Write a function data to the
designated code
C
F
This command is available when
“opErAte”
is displayed in the set point editing mode.
Summary of Contents for Explosion Protected Weighing Indicator AD-4403-FP ST Series
Page 22: ...2 7 Blank page...
Page 36: ...6 3 Blank page...
Page 38: ...6 5...
Page 40: ...6 7...
Page 53: ...6 20 Blank page...
Page 73: ...8 18 Blank page...
Page 83: ...Dimensions OP 20 Stand A 5...
Page 85: ...A 7 Blank page...
Page 87: ...B 2 Blank page...
Page 105: ...Fig Apx D 02 Gravity map WMF D 2...
Page 107: ...Form F Factory setting rSF 02 5 Command mode Address None rSF 09 0 Address 1 rSF 09 1 E 2...