2-2
2-2 Wiring
Open the front panel door removing screws on the front panel, and connect a power cable, load cell
cables, and control signal cables to the terminals inside the case. Follow the instructions below.
Cable used
Load cell
The use of a 6 wire shielded cable is recommended in order to reduce
weighing error. (When using a 4-wire cable, connect pin # 34 to # 35, and
pin # 36 to # 37.)
If the system requires two or more load cells, use an explosion-protected
type of summing box to input the signal into the indicator.
Control signal cable
Use a shielded cable and connect its shield to pin # 40 (Frame ground).
Power cable
Terminal
Description
E Ground
N
L
AC power supply
Load cell cable
Terminal
Description
Cable color
(ST Series only)
34 EXC+
Excitation
+
Red
35 SEN+
Sense
+
Orange
36 SEN-
Sense
-
Black
37 EXC-
Excitation
-
White
38 SIG+
Signal
+
Green
39 SIG-
Signal
-
Blue
40
F.G.
Frame ground
Yellow
Summary of Contents for Explosion Protected Weighing Indicator AD-4403-FP ST Series
Page 22: ...2 7 Blank page...
Page 36: ...6 3 Blank page...
Page 38: ...6 5...
Page 40: ...6 7...
Page 53: ...6 20 Blank page...
Page 73: ...8 18 Blank page...
Page 83: ...Dimensions OP 20 Stand A 5...
Page 85: ...A 7 Blank page...
Page 87: ...B 2 Blank page...
Page 105: ...Fig Apx D 02 Gravity map WMF D 2...
Page 107: ...Form F Factory setting rSF 02 5 Command mode Address None rSF 09 0 Address 1 rSF 09 1 E 2...