6-18
6-11 CALF-14=8 Weighing Mode
Check weighing 3)
Output signal
Output condition
Zero band
Gross weight
≤
Zero band
Hi-Hi Hi-Hi
limit
<
Net weight
Hi Hi
limit
<
Net weight
Go Lo
limit
≤
Net weight
≤
Hi limit
Lo
Net weight
< Lo
limit
Lo-Lo
Net weight
< Lo-Lo
limit
•
When an output condition is established, a relevant output terminal is turned on (power continuity with the output
common) or off. “Full flow”, “Medium flow”, and ”Dribble flow”: OFF, the other outputs: ON.
•
Select an output terminal at the outF-01-06 setting. See {Appendix C: Function List}.
•
Signal is output only in the normal mode. No output in the calibration mode nor function setting mode.