![AND Explosion Protected Weighing Indicator AD-4403-FP ST Series Instruction Manual Download Page 75](http://html1.mh-extra.com/html/and/explosion-protected-weighing-indicator-ad-4403-fp-st-series/explosion-protected-weighing-indicator-ad-4403-fp-st-series_instruction-manual_2941315075.webp)
9-2
Each check item
Display
Check Item
“Chc PrG”
Program Version
This check displays the version of the program stored in the ROM. The version number is displayed as
“vEr X.XX”
.
*
Press [ESC] to return to the check item-selecting mode.
“Chc Ad”
A/D Converter
The load cell Input voltage is displayed in the unit of mV/V. Applying offset voltage to the load cell can
check the A/D Converter function.
Keep the record of the input voltage for future maintenance work.
Apply some voltage with [F] and [ENTER] . “
□
”should move upward when applied,.
*
Press [ESC] to return to the check item-selecting mode.
# Problem
Cause
1
With +SIG and –SIG Load Cell terminals shorted,
0.0 mV/V is not displayed.
•
Incorrect wiring
•
Not sufficient input voltage
•
Damaged A/D converter
2
When not loaded, the displayed value is out of a
range of 0.0 mV/V to 2.0 mV/V.
•
Incorrect rate of load cell
•
Damaged load cell
3
When loaded, the displayed value is smaller than
that of # 2, or larger than 3.2 mV/V.
•
Incorrect wiring
•
Incorrect rate of load cell
4
“
□
” moves upward when [F] or [ENTER] pressed.
However no increase in value on the display.
•
Incorrect wiring
“Chc KEy”
Key Switches
Key number assignment
Front panel key
Pressing a key shows a relevant number (see the key
number assignment on the left.) The number is
displayed as
“no.XX”
.
([FUNC.] displays
“no.12”
)
Left keypad is of type A model. However key number
assignment is same between type A and B.
Calibration disable switch
Switching the calibration disable switch to the ON
position or OFF.
ON:
“S1”
OFF:
“ ”
(blank)
*
To return to the check item-selecting mode, press
[ESC] a little longer (more than 0.3 seconds)
*
When no switch is pressed,
“OFF”
is displayed.
*
If two or more switches are pressed,
“Err”
is
displayed.
Summary of Contents for Explosion Protected Weighing Indicator AD-4403-FP ST Series
Page 22: ...2 7 Blank page...
Page 36: ...6 3 Blank page...
Page 38: ...6 5...
Page 40: ...6 7...
Page 53: ...6 20 Blank page...
Page 73: ...8 18 Blank page...
Page 83: ...Dimensions OP 20 Stand A 5...
Page 85: ...A 7 Blank page...
Page 87: ...B 2 Blank page...
Page 105: ...Fig Apx D 02 Gravity map WMF D 2...
Page 107: ...Form F Factory setting rSF 02 5 Command mode Address None rSF 09 0 Address 1 rSF 09 1 E 2...