8-8
8-5 Data Transmitting Mode
OP-03/04 covers all transmitting modes. However standard serial output does not.
See the table below.
Stream
The data is transmitted in synchronization with the update of the display.
When the display rewrite exceeds the baud rate, data transmission is
suspended until the next display update.
The transmitted data is the same as what is being displayed. Therefore,
non-displayed data is never transmitted.
Auto print
The data is transmitted each time weighing has been completed. The
weighing completion timing differs depending on the weighing mode.
See {6. Weighing}.
*
To perform automatic accumulation and Auto Print at the same time
at batch finish, set functions as below.
•
Sq F-02(Automatic accumulation): “1” or “2”
•
SiF-02 (Data transmitting mode):
“
2
”
(Auto print)
•
rSF-02 (OP-03/04): “2” (Auto print)
Manual print
The data is transmitted when the “print command for manual print” is
sent with the Control input.
Standard serial output
Accumulation
timing print
The data is transmitted when [F] is pressed and the “print command for
accumulation” is sent.
Command
This mode analyzes the “command” received from the host computer,
etc. provides relevant processing, and responds with a result. For
example, if an “MT” command is received, the tare will be subtracted and
its result (whether tare subtraction is allowed or not) will be sent as a
response.
The address capability is available only in this mode.
OP-03:RS-422/485, OP-04:RS-232C
Output per Sampling
There are two types: gross and net. This mode outputs the set point
state as well as weights. This is a stream operation outputting at a rate of
100 times/second in synchronization with sampling.
Summary of Contents for Explosion Protected Weighing Indicator AD-4403-FP ST Series
Page 22: ...2 7 Blank page...
Page 36: ...6 3 Blank page...
Page 38: ...6 5...
Page 40: ...6 7...
Page 53: ...6 20 Blank page...
Page 73: ...8 18 Blank page...
Page 83: ...Dimensions OP 20 Stand A 5...
Page 85: ...A 7 Blank page...
Page 87: ...B 2 Blank page...
Page 105: ...Fig Apx D 02 Gravity map WMF D 2...
Page 107: ...Form F Factory setting rSF 02 5 Command mode Address None rSF 09 0 Address 1 rSF 09 1 E 2...