4-2
Actual load calibration procedures
Calibration procedures vary by the type of calibration. Follow the step number in the table below to
calibrate.
* When
“C ErrXX”
is displayed, there is something wrong with the calibration. See {Appendix
B: Error Codes}.
Zero calibration only
Span calibration only
Zero and Span calibration
Get in the calibration mode.
1 1 1
Press [CALF] while pressing [ENTER] in the normal mode, and press [ENTER] next.
“CAL Set”
is displayed. (ready to calibrate)
*
Press [ESC] to return to the normal mode.
Calibrate Zero.
2 2
Press [ENTER] to get into the zero calibration mode.
Main display:
“CAL 0”
/ Sub display:
“
”
*
To browse the gross weight, press [SET POINT], it will be displayed on the sub
display. Press [SET POINT] to erase this display.
*
Press [ESC] to return to the normal mode.
Press [ENTER] at a stable reading (
“-”
is on beneath “Stable”) with no load.
“-------”
is displayed for 2 seconds and the displays show;
Main display:
“CAL SPn”
/ Sub display:
“
Capacity weight
”
*
To end calibration without zero calibrated, press [ESC],
“CAL End”
is displayed.
Move on to Span calibration.
2
To skip zero calibration and perform span calibration only, press [ENTER]. When
“CAL 0”
is displayed, press [F].
Calibrate Span.
3 3
The displays show;
Main display:
“CAL SPn”
/ Sub display:
“
Capacity weight
”
*
The capacity weight is of the value set in CALF-04.
*
To browse the gross weight, press [SET POINT], it will be displayed on the sub
display. Press [SET POINT] to erase this display.
*
Press [ESC] to return to the normal mode.
*
To end calibration without span calibrated, press [ESC],
“CAL End”
is displayed.
Input the weight of the calibration mass using the numerical keys of ([0] - [9]).
Press [ENTER] at a stable reading (
“-”
is on beneath “Stable”).
“-------”
is displayed for 2 seconds and the displays show;
Main display:
“CAL End”
/ Sub display:
“
”
End Calibration after saving the calibration data in EEPROM.
3 4 4
Press [ENTER] to save the calibration data in the EEPROM and return to the normal
mode.
*
To return to the normal mode without saving the calibration data, press [ESC].
Summary of Contents for Explosion Protected Weighing Indicator AD-4403-FP ST Series
Page 22: ...2 7 Blank page...
Page 36: ...6 3 Blank page...
Page 38: ...6 5...
Page 40: ...6 7...
Page 53: ...6 20 Blank page...
Page 73: ...8 18 Blank page...
Page 83: ...Dimensions OP 20 Stand A 5...
Page 85: ...A 7 Blank page...
Page 87: ...B 2 Blank page...
Page 105: ...Fig Apx D 02 Gravity map WMF D 2...
Page 107: ...Form F Factory setting rSF 02 5 Command mode Address None rSF 09 0 Address 1 rSF 09 1 E 2...