![AND Explosion Protected Weighing Indicator AD-4403-FP ST Series Instruction Manual Download Page 50](http://html1.mh-extra.com/html/and/explosion-protected-weighing-indicator-ad-4403-fp-st-series/explosion-protected-weighing-indicator-ad-4403-fp-st-series_instruction-manual_2941315050.webp)
6-17
6-10 CALF-14=7 Weighing Mode
(Check weighing 2)
Output signal
Output condition
Zero band
Gross weight
≤
Zero band
Hi-Hi Target
weight
+
Hi-Hi limit
<
Net weight
Hi
Target weight +
Hi limit
<
Net weight
≤
Target weight +
Hi-Hi limit
Go
Target weight -
Lo limit
≤
Net weight
≤
Target weight + Hi limit
Lo
Target weight - Lo-Lo limit
≤
Net weight
<
Target weight - Lo limit
Lo-Lo
Net weight
<
Target weight - Lo-Lo limit
•
When an output condition is established, a relevant output terminal is turned on (power continuity with the output
common) or off. “Full flow”, “Medium flow”, and ”Dribble flow”: OFF, the other outputs: ON.
•
Select an output terminal at the outF-01-06 setting. See {Appendix C: Function List}
•
Signal is output only in the normal mode. No output in the calibration mode nor function setting mode.
Summary of Contents for Explosion Protected Weighing Indicator AD-4403-FP ST Series
Page 22: ...2 7 Blank page...
Page 36: ...6 3 Blank page...
Page 38: ...6 5...
Page 40: ...6 7...
Page 53: ...6 20 Blank page...
Page 73: ...8 18 Blank page...
Page 83: ...Dimensions OP 20 Stand A 5...
Page 85: ...A 7 Blank page...
Page 87: ...B 2 Blank page...
Page 105: ...Fig Apx D 02 Gravity map WMF D 2...
Page 107: ...Form F Factory setting rSF 02 5 Command mode Address None rSF 09 0 Address 1 rSF 09 1 E 2...