Specifications | 2-5
Calibration Gas Requirements for Hot-Wire Type Combustibles Detector
Use calibration gases @ 10 PSIG, 1.5 SCFH (0.70 kg/cm
2
, 0.7 L/min.)
o
2
span gas:
Air or from 1.0 to 100% o
2
, balance n
2
o
2
comb &
CH4 Zero Gases:
From 0.1 to 10% o
2
, balance n
2
comb span gas:
60 to 80% (ppm ranges) or 40 to 60% (% ranges) of the selected
comb. recorder output range in certified mixtures of
CO + H
2
, 3-4% o
2
, balance n
2
.
CH
4
span gas
2% CH
4
, 8% to 10% o
2
, balance n
2
Calibration Gas Requirements for RTD Type Combustibles Detector
Use calibration gases @ 10 PSIG, 1.5 SCFH (0.70 kg/cm
2
, 0.7 L/min.)
o
2
span gas:
Air or from 1.0 to 100% o
2
, balance n
2
Comb & CH4
span gases:
From 0.1 to 10% o
2
, balance n
2
comb span gas:
use a combined combustibles and methane span gas containing 60 to
80% of the full scale combustibles operating range in certified mixtures
of CO (or CO + H
2
), 2% CH
4
, 8 to 10% o
2
, balance n
2.
System Compliance
EMc compliance:
Safety Compliance:
2004/108/Ec
73/23/EEc
NOTES
:
1.
All static performance characteristics are with operating variables constant.
2.
System accuracy referenced to 0.1 to 10% calibrated range.
3.
Response is to calibration gas (without flame arrestors).
Summary of Contents for Thermox Series 2000 WDG-IV
Page 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Page 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Page 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Page 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Page 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Page 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Page 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...