3-20 | Thermox Series 2000 WDG-IV / IVC / IVCM
This wiring section shows you how to make the following connections:
•
AC mains supply wiring to control unit
•
AC mains supply wiring to sensor
•
Control unit to sensor
•
Control unit to remote calibration unit
•
Control unit to alarm devices
•
Control unit to current output devices
•
Control unit to host computer (RS-485 communications)
In addition, this wiring section provides mandatory EMC grounding,
shielding, and noise protection requirements.
If you are considering adding option cards at a later time, such as a
combustibles card, you may want to connect the required intercon-
necting wires when you first install your analyzer, especially if run-
ning the wires over long distances through conduit.
General Wiring and Conduit Requirements
This section describes general wiring and conduit requirements.
•
Sensor wiring conductors must be rated at a minimum of 80 °C. All
other wiring conductor ratings should be for the minimum tempera-
ture required for the equipment being connected to the analyzer, but
not less than 60 °C.
•
Use only the applicable NEMA-approved conduit fittings or cable
fittings to maintain the NEMA rating for the Series 2000 controller
or sensor enclosure. If not using a conduit entry, leave the factory
NEMA-approved plugs intact. Never leave any holes unplugged.
•
Follow all applicable electrical codes for your location.
•
Follow proper grounding, shielding and noise protection practices as
described in this section.
•
Use twisted-pair cable with an overall braided shield, or twisted-pair
cable in rigid metal conduit for all sensor and signal wiring.
NOTE
Summary of Contents for Thermox Series 2000 WDG-IV
Page 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Page 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Page 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Page 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Page 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Page 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Page 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...