7-12 | Thermox Series 2000 WDG-IV / IVC / IVCM
RTD Combustibles Detector Replacement
Use Figure 7-6:
1. Remove the detector leads from terminal strips E, F, H, and I on the
sensor board.
2. Loosen the compression fitting nut that holds the combustible detec-
tor in place and remove the detector from the combustible detector
block.
3. It is very important that you orient the new combustibles detector cor-
rectly inside the combustibles detector block (sample flow path must
contact both detector leads at the same time). To do this, ensure that
the flat part of the combustibles detector is facing down (see Figure 7-6
for correct orientation).
4. Install the new combustibles detector with nut and ferrules. Be ex-
tremely careful; do not damage the detector elements when install-
ing. Ensure the proper orientation of the combustibles detector in the
combustibles detector block.
5. Reconnect the detector leads to the sensor board by reversing the ac-
tions taken in Step 1 above.
Furnace Assembly
Combustibles
Detector Block
Ferrule
Detector Head
Compression
Fitting Nut
Side View
Front View
Sample
flow
flow
Flat must
face down
Figure 7-6. Combustibles detector replacement (RTD).
Summary of Contents for Thermox Series 2000 WDG-IV
Page 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Page 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Page 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Page 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Page 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Page 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Page 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...