A-16 | Thermox Series 2000 WDG-IV /I VC / IVCM
45/69
Cal_h10
Auto calibration hour value
R/W
46/70
Cal_h1
Auto calibration hour value
R/W
47/71
Cal_m10
Auto calibration minute value
R/W
48/72
Cal_m1
Auto calibration minute value
R/W
49/73
Ver_h10
Auto verify time hour value
R/W
4A/74
Ver_h1
Auto verify time hour value
R/W
4B/75
Ver_m10
Auto verify time minute value
R/W
4C/76
Ver_m1
Auto verify time minute value
R/W
4D/77
Pressure
Process pressure.
R/W
4E/78
Optemp
The cell temperature setpoint
Read only
4F/79
Optemp_range
Operating temperature for sensor, 0 is
normal, 1 is high, 2 is RM CEM
IQ
R/W
50/80
Cmbref
Reference voltage reading for Combustibles
channel
Read only
51/81
Cmbact
Active voltage reading for Combustibles
channel
Read only
52/82
Cmbgnd
Ground reading for Combustibles channel
Read only
53/83
Mthref
Reference voltage reading for the Methane
channel.
Read only
54/84
Mthact
Active voltage reading for Methane channel
Read only
55/85
Mthgnd
Ground reading for Methane channel
Read only
56/86
AT
Cell slope
Read only
57/87
AT_ratio
Cell slope used during last calibration.
Read only
58/88
Tc_ratio
Cell/thermocouple compensation multiplier
Read only
59/89
Cmbslope
Combustibles slope during last calibration
Read only
5A/90
Mthslope
Methane slope during last calibration
Read only
5B/91
Cal_cycle
Interval between auto cals [1 hour to 90
days]
R/W
5C/92
Ver_cycle
Interval between auto verifies [1 hour to
90 days]
R/W
Summary of Contents for Thermox Series 2000 WDG-IV
Page 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Page 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Page 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Page 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Page 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Page 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Page 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...