5-22 | Thermox Series 2000 WDG-IV / IVC / IVCM
Set Cal timers - Set time and Date
Set the date or time for the internal control unit clock.
Make sure the current time and date are correct before setting any
auto cal/auto verify parameters.
1. From the
Set Cal Timers
menu, select
Set Time & Date
.
You are prompted to enter the new time:
The time is: YY:YY
Enter new time Hr:Mn
2. Enter the new time, then press
Enter
. As you type the new time, it
overwrites the Hr: Mn text. The Series 2000 control unit uses military
time so 16:00 equals 4:00 p.m., and 04:00 equals 4:00 a.m. If the current
time is correct, press
Enter
without changing the value.
You are then prompted to enter the correct date:
The date is: YY/XX/ZZ
Enter date: Mo/Da/Yr
3. As you type the new date, it overwrites the mo/da/yr text on the
display. Enter the month, the day, the year, then press
Enter
. If the
current date is correct, press
Enter
without changing the value.
Set Cal Timers - Auto Cal On/Off
Disable or re-enable the auto calibration or auto-verification cycle. If you
know your process will be down for an extended period of time, you may
want to temporarily disable the automatic calibration or verification cycle.
When the process is back on-line, you can re-enable the auto calibration or
verification cycle and retain all your timer settings.
1. From the
Set Cal Timers
menu, select
Auto Cal On/Off
.
2. Select
Auto Cal On/Off
or select
Auto Verify On/Of
f, depending on
whether you want to enable or disable the auto verify or auto calibrate
cycle.
3. Select
Disable Auto Cal
(or
Disable Auto Verif
y) to disable the auto
calibration or auto verification cycle. Select
Enable Auto Cal
(or
En-
able Auto Verif
y) to enable the auto calibration or verification cycle.
NOTE
Summary of Contents for Thermox Series 2000 WDG-IV
Page 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Page 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Page 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Page 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Page 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Page 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Page 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...