Serial Communications
|
A-21
6E/110
Dig_Input
Status of digital inputs(1=closed/on,
0=open/off)
bit 0
– CAL 0
IQ
bit 0 –
Dig In
S2
bit 1
– CAL 1
IQ
Read only
6F/111
70/112
SysTst_Flag
System test that was preformed
1-Calibrate A/D
2-Ram Test
4-Test EEPROM
7-Erase Ram
8-Erase EEPROM
9- Test Watchdog
R/W
71/113
Passw
Current password
R/W
72/114
BoxCtl
IQ
Value that corresponds to the box set point
0 – 230 C
1 – 270 C
2 – 360 C
3 – Full on control
R/W
73/115
InjectGas
Turn on gas for inject cal gas menu
0 - Span gas
1 - Zero gas
2 - Comb gas
3 - Methane gas (if 3 span value RCU
installed)
4 - Process gas
R/W
74/116
Node_Addr
The current node address
R/W
75/117
Read only
76/118
Cmbsens
Combust sensitivity associated with the
last calibration
Read only
77/119
Mthsens
Methane sensitivity associated with the
last calibration
Read only
78/120
AFRatio
The calculated A/F ratio
Read only
79/121
AFRatio_CalSP
The zero cal gas A/F ratio set point
R/W
7A/122
Wobbe
The calculated WOBBE value
Read only
7B/123
Wobbe_SP
The WOBBE Factor
R/W
7C/124
Gas4_time
Amount of time the methane span gas will
flow
IQ
(in seconds)
R/W
7D/125
AFRatio_CalSP2
The A/F Ratio set point at low end
R/W
7E/126
XSO2Low_SP
The XS O2 set point at low end
R/W
7F/127
SW Version
Current software version #
Read only
Summary of Contents for Thermox Series 2000 WDG-IV
Page 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Page 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Page 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Page 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Page 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Page 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Page 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...