5-16 | Thermox Series 2000 WDG-IV / IVC / IVCM
Cal Gas Value
Enter span and zero calibration gas values to match the calibration gas
cylinders you use to calibrate or verify your system. These values will be
used for all automatic, remote, or manual calibrations and verifications.
•
O
2
span gas - high calibration gas
•
Zero gas - low calibration gas
•
Combustibles span gas (optional)
•
Methane span gas (optional)
The O
2
span gas value you enter must be higher than the O
2
zero gas
value.
1. From the
Calibrate
key menu, select
Cal Gas Value
.
You are prompted to enter the span gas value:
Span Gas = xx.xx% O2
New Value?
2. Enter the span gas value, then press
Enter
. You can also include a
decimal point (for example, 20.9% O
2
). If you want to keep the current
span gas value unchanged, press
Enter
without entering a value.
You are prompted to enter the zero gas value:
Zero Gas = xx.xx% O2
New Value?
3. Enter the zero gas value, then press
Enter
. You can enter up to two
decimal points (for example, 2.00% O
2
). If you want to keep the cur-
rent zero gas value unchanged, press
Enter
without entering a value.
Combustibles option
If you have the combustibles option, you are then prompted to enter
the combustibles span gas value:
Comb Span = xxxx YY
New Value?
NOTE
Summary of Contents for Thermox Series 2000 WDG-IV
Page 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Page 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Page 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Page 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Page 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Page 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Page 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...