Calibration | 5-11
If you have the methane option, you are also prompted to apply the
methane span gas to the calibration gas inlet port of the sensor (tubing
must be free of oil and dirt):
Inject Meth Span Gas
ENTER to Continue
CANCEL to Abort
4.
Press
Enter
when you have applied your methane span gas. You are
prompted to calibrate the methane span gas.
O2 xx.x %\Cmb yyyy Set yyyy
CH4 zzzz Set zzzz
Press ENTER to Span Meth
5.
Press
Enter
when the methane span gas reading stabilizes on the far left
of the display.
If you specified a recovery time, the system will display the recovery time
remaining:
Recovery xx:xx
System Calibrating
The system then returns to monitoring process readings.
Be sure to reinsert the 1/4” tube cap on the calibration inlet port on
the sensor once you have completed the manual calibration. This pre-
vents air from getting into the sample gas line during normal opera-
tions.
NOTE
Summary of Contents for Thermox Series 2000 WDG-IV
Page 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Page 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Page 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Page 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Page 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Page 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Page 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...