Overview
|
1-3
Figure 1-1. Zirconium oxide cell principle of operation.
Since the voltage of the cell is temperature dependent, the cell is main-
tained at a constant temperature. Some newer high temperature insitu
models use the heat from the process to heat the sensor, and the process
temperature is continuously measured and used in the software calcula-
tion. The oxygen content is then determined from the Nernst equation:
E = In
RT
4F
O
1
O
2
where R and F are constants, T is absolute temperature, and O
1
and O
2
are
the oxygen partial pressures on either side of the cell.
For measuring oxygen in non-combustibles gases, the calibration of an
analyzer is obtained from the formula:
AT = 48.0 at 695°C
E = A*T*Log
20.9%
O
2
Unk%
Where
A
is a constant,
T
is the cell temperature on an absolute scale (°C +
273) and O
2
Unk% is the unknown oxygen concentration of the gas to be
analyzed, and which is calculated by the analyzer.
Summary of Contents for Thermox Series 2000 WDG-IV
Page 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Page 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Page 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Page 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Page 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Page 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Page 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...