5-18 | Thermox Series 2000 WDG-IV / IVC / IVCM
Inject Cal Gas
Manually adjust the pressure and flow of your calibration gases (pressure
of calibration gas cylinders and flow into sensor) by cycling through the
solenoid valves on the RCU. No calibration or verification data is collected
while in this menu option. See the “Mechanical Installation” section in
Chapter 2 for recommended calibration gas pressure and flow require-
ments.
This menu option requires that an auto cal card and RCU are in-
stalled.
1. Connect the span and zero gas cylinders to the RCU.
2. From the
Calibrate
key menu, select the
Inject Cal Gas
menu option.
The following message will appear on your display, indicating that the
RCU has introduced the span gas to the sensor:
Span gas is flowing
You will be prompted to press
Enter
when you have manually set the
pressure and flow for the span gas:
ENTER for next Gas.
CANCEL to quit.
3. When you have adjusted the pressure and flow for the span gas to the
desired levels, press
Enter
.
The following message will appear on your display, indicating that the
RCU has now switched the solenoid valves and has introduced the zero
gas to the sensor:
Zero gas is flowing
4. Adjust the pressure and flow for the zero gas to the desired levels. If
you don’t have the combustibles option, press the
Cancel
key to end
the cycle.
CANCEL to quit
Then, if you specified a recovery time, the system will display the recov-
ery time remaining:
Recovery xx:xx
NOTE
Summary of Contents for Thermox Series 2000 WDG-IV
Page 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Page 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Page 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Page 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Page 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Page 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Page 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...