Controller / User Interface | 4-11
Setting up a primary calibration
Make sure you have properly entered calibration gas values using the
Cal
Gas Values
option from the
Calibrate
key before starting your primary
calibration.
1. From the
Setup
key, select
Primary Calib
.
2. If you have the auto calibration card option (and a remote calibra-
tion unit), you will be prompted to select
Auto Calibrate
,
Remote
Calibrate
, or
Manual Calibrate
. If you don’t have the auto calibration
option, the system defaults to a manual calibration, and you will be
prompted to inject span gas into the calibration gas inlet port.
From this point, operations continue as they do for a regular calibration.
Please see the “Initiate Cal/Initiate Verify” section in Chapter 5 for help
on how to perform the calibration.
If you have the combustibles option, during a primary calibration the
system will not use the combustible span gas to calibrate the system.
When the primary calibration is finished, the system will notify you
that a standard calibration is required by displaying a
Calibration
Required
message.
Oxygen readings on the display will not be accurate until the regular
calibration is performed.
A 15-minute recovery time follows the primary calibration to allow
the cell to stabilize. Perform the regular calibration after this 15 minute
stabilization period,
During the recovery time immediately following the primary calibra-
tion, you may see some temperature related error messages. You can
ignore these messages and assume no problem exists. This is because
the cell temperature is adjusted after a primary calibration.
NOTE
NOTE
Summary of Contents for Thermox Series 2000 WDG-IV
Page 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Page 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Page 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Page 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Page 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Page 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Page 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...