5-24 | Thermox Series 2000 WDG-IV / IVC / IVCM
Combustibles and Methane options
You will be prompted to enter the new combustibles span gas duration time:
Comb Span Time = xx:yy
Enter new time: Mn:Sc
Enter the new combustibles span gas duration time, then press
Enter
. You
can set this calibration gas time to 0 if you don’t want to include it in your
automatic calibration sequence. If you want to use the currently displayed
time, press
Enter
without entering a new value.
You will be prompted to enter the new methane span gas duration time:
Meth Span Time = xx:yy
Enter new time: Mn:Sc
Enter the new methane span gas duration time, then press
Enter
. You
can set this methane gas time to 0 if you don’t want to include it in your
automatic calibration sequence. If you want to use the currently displayed
time, press
Enter
without entering a new value.
Set Cal Timers - Recovery Duration
Define a recovery time so the control unit can return to reading process
gases after reading calibration gases without affecting the analog outputs
or the triggering of alarms.
This menu option also can be used to prevent triggering of alarms if, for
example, your low alarm is set to activate if readings fall below 3%, and
your calibration gas is 2%.
This menu option applies to all automatic, remote, and manual cali-
brations and verifications.
1. From the
Set Cal Timers
menu, select
Recovery Duration
.
You will be prompted to enter a calibration recovery duration:
Cal recovery = xx:yy
Enter new time: Mn:Sc
2. Enter the recovery time, then press
Enter
. If you want to use the cur-
rently displayed recovery time, press
Enter
without entering a new
value.
NOTE
Summary of Contents for Thermox Series 2000 WDG-IV
Page 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Page 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Page 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Page 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Page 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Page 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Page 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...