Controller / User Interface | 4-23
3. Enter the Alarm 4 set point value.
If you don’t have the combustibles option installed, you are done (press
Cancel
to exit this menu option).
Combustibles Option Alarms
If you do have the combustibles option installed, you are prompted to
enter the Alarm 5 set point value:
Alarm #5= xx yy
New Value?
where xx is the current value and yy is either PPM combustibles or %
combustibles, depending on the scale you are using.
4. Enter the Alarm 5 set point value. Note that combustibles alarms are
always high alarms (can’t be set up as a low alarm).
You are prompted to enter the Alarm 6 set point value:
Alarm #6= xx yy
New Value?
where xx is the current value and yy is either PPM combustibles or %
combustibles, depending on the combustibles scale you are using.
5. Enter the Alarm 6 set point value. Note that combustibles alarms are
always high alarms. So, for example, Alarm 5 can be high and Alarm 6
can be high/high.
If you don’t have the methane option, you are done (press
Cancel
key or
Enter
to exit this menu option).
Methane Option Alarms
If you do have a methane detector, you are then prompted to enter the
Alarm 7 set point value:
Alarm #7=xx.x % CH4
New Value?
6. Enter the Alarm 7 set point value. Note that methane alarms are al-
ways high alarms.
Summary of Contents for Thermox Series 2000 WDG-IV
Page 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Page 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Page 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Page 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Page 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Page 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Page 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Page 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...