8-10 | 3050 RM Moisture Analyzer
Dryers / Contaminant Traps
Removing the dryers
1. At the rear of the analyzer, locate the dryer compartment on the right-
hand side (Figure 8-8).
2. Use a 7/64” hex wrench to loosen the four captured screws on the
compartment cover.
3. Remove the cover from the analyzer.
4. The single dryer is located in the bottom position. If your analyzer is
equipped with two dryers or a dryer and a trap there will be a dryer
in bottom position and a dryer or a trap in the top position.
5. Loosen the 1/8 inch fittings on the dryer(s)/trap and remove the
dryer(s)/trap from the analyzer. Discard the used gaskets from the fit-
tings.
Figure 8-8.
Model 3050 RM.
Dryers
Summary of Contents for 3050-RM