User Interface
|
4-9
In practice, some empirical modification is required; this helps to compen-
sate for the non-ideal performance of the meter. For example, the actual
factor for argon is 1.395, rather than the 1.401 estimated in the example.
Please consult AMETEK Process Instruments for assistance, before run-
ning gases that are not listed in the menu on the 3050-RM analyzer.
If a custom gas is entered, the conversion to units of dew point will be
based on an ideal gas model, without any enhancement factor.
Gas Saver
In the default configuration the Model 3050-RM analyzer runs with the
internal bypass valve open, so that the total flow through the analyzer
is greater than 1 SLPM. This high sample flow rate helps to improve the
response speed of the system. The gas-saver feature allows the user to
control the state of the bypass valve so that the analyzer runs on a sample
flow rate of 150 SCCM. The operation of the valve can be controlled
by checking the Gas Saver box (to close the bypass valve) or leaving it
unchecked(to leave the bypass valve open).
Sensor Saver
In the default configuration the Model 3050-RM analyzer operates with
a rapid cycle time to minimize system response time. The sensor-saver
feature allows the analyzer to operate with a slow cycle time, maximizing
the cell life but minimizing system response time. The operation of the
valve can be controlled by checking the Sensor Saver box (to operate with
a slow cycle time) or leaving it unchecked (to operate with a rapid cycle
time).
NOTE
Summary of Contents for 3050-RM
Page 11: ... xi ...
Page 20: ...1 8 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 24: ...2 4 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 27: ...Installation Start Up 3 3 Figures 3 1c Model 3050 RM dimensions top and sides ...
Page 44: ...4 2 3050 RM Moisture Analyzer Figure 4 2 Menu map for 3050 RM Moisture Analyzer ...
Page 63: ...Serial Communications 5 7 Figure 5 6 Defined Failure Codes ...
Page 64: ...5 8 3050 RM Moisture Analyzer Figure 5 7a Variable Table ...
Page 65: ...Serial Communications 5 9 Figure 5 7b Variable Table ...
Page 66: ...5 10 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 78: ...7 2 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 91: ...Service Parts 8 13 Figure 8 9 Electronic Module ...
Page 98: ...PPmV Centigrade Fahrenheit lbs mnscf mg Nm3 ppmW ...