5-4 | 3050-RM Moisture Analyzer
Echo (A)
Echo implements a loop back function to test the integrity of the serial
communication. Any text transmitted by the master is returned.
Example
:
Master:
>99ABCDEFG4E[CR]
Slave:
AABCDEFG1D[CR]
Bad Command (B)
This command is not recognized by the slave. It will therefore result in a
type 01 failure response.
Master:
>99BB4[CR]
Slave:
N01[CR]
Acknowledge (C )
Requests a simple acknowledge from the slave. Useful to determine if
there is a device at that node address.
Master:
>99CB5[CR]
Slave:
A[CR]
Read Data (F)
The data field of this frame contains the handle for a system variable. See
Figures 5-7a and 5-7b for a list of defined handles for system variables.
Example
:
Master:
>34F0512[CR]
Slave:
A1234540[CR]
Start Verify (G)
Initiates a verify cycle.
Example:
Master:
>34G??[CR] (Note the use of wildcard in place of checksum)
Slave:
A[CR]
Summary of Contents for 3050-RM
Page 11: ... xi ...
Page 20: ...1 8 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 24: ...2 4 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 27: ...Installation Start Up 3 3 Figures 3 1c Model 3050 RM dimensions top and sides ...
Page 44: ...4 2 3050 RM Moisture Analyzer Figure 4 2 Menu map for 3050 RM Moisture Analyzer ...
Page 63: ...Serial Communications 5 7 Figure 5 6 Defined Failure Codes ...
Page 64: ...5 8 3050 RM Moisture Analyzer Figure 5 7a Variable Table ...
Page 65: ...Serial Communications 5 9 Figure 5 7b Variable Table ...
Page 66: ...5 10 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 78: ...7 2 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 91: ...Service Parts 8 13 Figure 8 9 Electronic Module ...
Page 98: ...PPmV Centigrade Fahrenheit lbs mnscf mg Nm3 ppmW ...