Installation & Start-Up
|
3-11
Electrical Connections (Figure 3-6)
This analyzer has an internal AC voltage selector that must be set
correctly to prevent damage. Only qualified service personnel may ac-
cess the interior of the analyzer. If qualified service personnel are not
available, contact AMETEK service.
Voltage selection
The 3050-RM analyzer comes from the factory set to operate on 120 VAC.
If you must switch the setting to 240 VAC (or back to the 120 VAC setting),
follow the procedure below.
You must follow this procedure to prevent damage to/malfunction of
the analyzer.
1. Ensure that the area is non-hazardous and the analyzer has been de-
energized.
2. Remove the terminal cover on back panel.
3. Select 115 or 230 VAC on the voltage selector (10) located on the top
left-hand side of the user interface board.
4. Replace terminal cover.
5. Clearly mark the operating voltage on the rear panel analyzer serial
number label in the lower right-hand corner.
!
WARNING
!
CAUTION
Figure 3-6.
115 / 230 voltage selection switch.
Summary of Contents for 3050-RM
Page 11: ... xi ...
Page 20: ...1 8 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 24: ...2 4 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 27: ...Installation Start Up 3 3 Figures 3 1c Model 3050 RM dimensions top and sides ...
Page 44: ...4 2 3050 RM Moisture Analyzer Figure 4 2 Menu map for 3050 RM Moisture Analyzer ...
Page 63: ...Serial Communications 5 7 Figure 5 6 Defined Failure Codes ...
Page 64: ...5 8 3050 RM Moisture Analyzer Figure 5 7a Variable Table ...
Page 65: ...Serial Communications 5 9 Figure 5 7b Variable Table ...
Page 66: ...5 10 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 78: ...7 2 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 91: ...Service Parts 8 13 Figure 8 9 Electronic Module ...
Page 98: ...PPmV Centigrade Fahrenheit lbs mnscf mg Nm3 ppmW ...