1-4 | 3050-RM Moisture Analyzer
Sample flow
Normal operation of the Model 3050-RM analyzer uses an internal bypass,
which increases the response speed of the system. However, the 3050-RM
is capable of running in a gas-saver mode as well, which allows the ana-
lyzer to run on a sample volume of approximately 150 SCCM.
There are four separate sample paths in the 3050-RM analyzer manifold.
The first is a simple bypass that allows for an increased flowrate of
sample gas. This means that when you open the bypass valve (SV1),
you reduce the time required to equilibrate all portions of the sample
system that exist upstream of the bypass valve. A capillary tube,
located immediately downstream of SV1 (the coil shape in Figure 1-2),
limits the flow rate to approximately 1 standard liter per minute (the
exact value of the bypass flow is a function of the inlet pressure to the
analyzer).
Figure 1-2.
Flow schematic for the 3050-RM
sample manifold.
Summary of Contents for 3050-RM
Page 11: ... xi ...
Page 20: ...1 8 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 24: ...2 4 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 27: ...Installation Start Up 3 3 Figures 3 1c Model 3050 RM dimensions top and sides ...
Page 44: ...4 2 3050 RM Moisture Analyzer Figure 4 2 Menu map for 3050 RM Moisture Analyzer ...
Page 63: ...Serial Communications 5 7 Figure 5 6 Defined Failure Codes ...
Page 64: ...5 8 3050 RM Moisture Analyzer Figure 5 7a Variable Table ...
Page 65: ...Serial Communications 5 9 Figure 5 7b Variable Table ...
Page 66: ...5 10 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 78: ...7 2 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 91: ...Service Parts 8 13 Figure 8 9 Electronic Module ...
Page 98: ...PPmV Centigrade Fahrenheit lbs mnscf mg Nm3 ppmW ...