Specifications
|
2-1
SPECIFICATIONS
All specifications subject to change without notice.
Compatible Gases
Air, argon, Air, argon, butane, carbon dioxide, ethane, ethene, Freon 12,
Freon 22, Freon 32, Freon 114, Freon 116, Freon 125, Freon 134A, Freon 152,
helium, hydrogren, krypton, natural gas, neon, nitrogen, oxygen, pro-
pane, propene, SF6, and xenon. Carbon dioxide requires a custom sensor
assembly. Contact the factory to confirm compatibility with other gases.
Read the Dryer Warning in the “Special Warnings and Information”
section in the beginning of this manual for more information.
Range
Calibrated range of 0 to 2500 PPMV.
Limit of Detection
0.1 PPMV nominal
Accuracy
± 0.1 PPMV or ±10% of reading, whichever is greater
Response Time
Typical 63% of a step change in either direction in less than 5 minutes
NOTE
!
CAUTION
Summary of Contents for 3050-RM
Page 11: ... xi ...
Page 20: ...1 8 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 24: ...2 4 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 27: ...Installation Start Up 3 3 Figures 3 1c Model 3050 RM dimensions top and sides ...
Page 44: ...4 2 3050 RM Moisture Analyzer Figure 4 2 Menu map for 3050 RM Moisture Analyzer ...
Page 63: ...Serial Communications 5 7 Figure 5 6 Defined Failure Codes ...
Page 64: ...5 8 3050 RM Moisture Analyzer Figure 5 7a Variable Table ...
Page 65: ...Serial Communications 5 9 Figure 5 7b Variable Table ...
Page 66: ...5 10 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 78: ...7 2 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 91: ...Service Parts 8 13 Figure 8 9 Electronic Module ...
Page 98: ...PPmV Centigrade Fahrenheit lbs mnscf mg Nm3 ppmW ...