3-18 | 3050-RM Moisture Analyzer
Initial Start-Up
1. Open the main sample shutoff valve at the source.
2. Adjust the sample pressure (20 to 50 PSIG).
Maximum inlet pressure is 50 PSIG. Damage can occur if higher
pressures are introduced to the analyzer.
3. Turn on the AC power at the source. Boot-up information will appear
on the display followed by the operation screen (Figure 3-14).
4. Once sample flow starts, the analyzer will be in the normal operating
mode. Lines 3 and 4 will flash various messages and alarm conditions
at start-up.
5. Allow the analyzer to warm up for a minimum of 45 minutes.
Dry-down period
Sample must be flowing during the dry-down period. You should allow 1
to 3 days for the analyzer to dry down and stabilize. During this period,
the concentration value shown on the display and the data recorded from
output should show a steadily decreasing trend. System alarm messages
may be displayed during this period.
When the dry down is complete, the concentration value on the display
should be stable and the recorded data should have leveled off.
If the concentration value is still decreasing after the initial dry down
period, allow additional dry down time.
The 3050-RM Moisture Analyzer is now fully operational with factory
default configuration of the analog output and concentration alarms.
NOTE
!
CAUTION
Figure 3-14.
Default normal operation screen.
Values are for example only.
Summary of Contents for 3050-RM
Page 11: ... xi ...
Page 20: ...1 8 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 24: ...2 4 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 27: ...Installation Start Up 3 3 Figures 3 1c Model 3050 RM dimensions top and sides ...
Page 44: ...4 2 3050 RM Moisture Analyzer Figure 4 2 Menu map for 3050 RM Moisture Analyzer ...
Page 63: ...Serial Communications 5 7 Figure 5 6 Defined Failure Codes ...
Page 64: ...5 8 3050 RM Moisture Analyzer Figure 5 7a Variable Table ...
Page 65: ...Serial Communications 5 9 Figure 5 7b Variable Table ...
Page 66: ...5 10 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 78: ...7 2 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 91: ...Service Parts 8 13 Figure 8 9 Electronic Module ...
Page 98: ...PPmV Centigrade Fahrenheit lbs mnscf mg Nm3 ppmW ...