2-2 | 3050-RM Moisture Analyzer
Sample Inlet Pressure Requirement
138 kPa to 345 kPa ( 20 PSIG to 50 PSIG )
Exhaust Pressure
0-103 kPa (0 to 15 PSIG)
Sample Flow Requirement
Approximately 150 SCCM in sample saver mode.
Approximately 1 SLPM with bypass open.
Inlet Gas Temperature
0 °C to 100 °C (32 °F to 212 °F)
Optimal results are obtained when inlet gas temperature is maintained at
60 °C (140 °F).
Outputs
4-line x 20-character LCD display
One 4 to 20 mA DC analog output
RS-232 / RS-485 serial port
Alarms/Alerts
Three independent contact closures of 30 VAC or 60 VDC maximum, 10 VA
or 1 A maximum, resistive for system alarm, moisture concentration alarm,
and data valid. All are fail-safe by default.
Alarms are available on RS-485 interface.
Environmental Conditions
5 °C to 50 °C (41 °F to 122 °F)
[Optimal performance in low-level applications is obtained when ambient tempera-
ture is maintained at a constant temperature (± 2 °C)]
90% RH maximum, noncondensing
IEC Pollution Degree 2
Maximum altitude 2,000 meters (6560 feet)
IEC Installation Category II
Indoor use only, IEC 529; IP40
Summary of Contents for 3050-RM
Page 11: ... xi ...
Page 20: ...1 8 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 24: ...2 4 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 27: ...Installation Start Up 3 3 Figures 3 1c Model 3050 RM dimensions top and sides ...
Page 44: ...4 2 3050 RM Moisture Analyzer Figure 4 2 Menu map for 3050 RM Moisture Analyzer ...
Page 63: ...Serial Communications 5 7 Figure 5 6 Defined Failure Codes ...
Page 64: ...5 8 3050 RM Moisture Analyzer Figure 5 7a Variable Table ...
Page 65: ...Serial Communications 5 9 Figure 5 7b Variable Table ...
Page 66: ...5 10 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 78: ...7 2 3050 RM Moisture Analyzer This page intentionally left blank ...
Page 91: ...Service Parts 8 13 Figure 8 9 Electronic Module ...
Page 98: ...PPmV Centigrade Fahrenheit lbs mnscf mg Nm3 ppmW ...