User Manual – Rev J
CS Series
California Instruments
67
ENTRY
DESCRIPTION
IMM
Immediate mode. The transient is started from
the front panel using the ENTER key.
BUS
Bus mode. The transient system is started by a
bus command or a group execute trigger (GET).
EXT
External mode. The transient system is started
by a user-provided external TTL trigger signal
on TRIGGER IN1.
TRAN STEP
Indicates the transient system execution mode. Available
modes are:
AUTO
When triggered, the transient system will
automatically execute each list point
sequentially without waiting for a trigger
between list points. This execution is paced by
the dwell time set for each data point.
ONCE
When triggered, the transient system will
execute the first list point and wait for a new
trigger once the dwell time expires. This allows
triggered execution of each step in the transient
list.
List parameters:
CURR
Step #
Current set point
CSLEW
Step #
Current slew rate in A/sec
FREQ
Step #
Frequency set point
FSLEW
Step #
Frequency slew rate in Hz/sec
FUNC
Step #
Waveform selection. Available choices are Sine,
Square, Clipped or any of the user provided
waveforms in waveform memory.
PAUSE Step
#
Delay
COMPV
Step #
Compliance voltage limit set point
DWELL
Step #
Dwell time in seconds.
Range is 0.001 to 900000
TTLT
Step #
ON:
Generates an output trigger pulse at
this list step.
OFF:
No output trigger.
The output trigger is available on the Function
Strobe SMC connector on the rear panel.