31
Power-On Sequence
Th e L4532A/ 34A may t ake up to a min u t e t o boot, du r in g which you will
see some p rogr ess messages on its disp lay. Then t he d isp lay p rovides
t hr ee sets of in for mation , in sequen ce:
ID and revision message
Fir st, t he in str umen t will pr ovide an ID an d
ver sion in fo (typ ical examp le shown ):
Th e fir st lin e disp lays t he mod el n u mber. Th e secon d lin e is in th e for m
f.ff-bb-gg-rr
, wher e:
f.ff
is th e in str umen t's in st alled fir mwar e ver sion .
bb
is t he FPGA ver sion for t he d igitizer 's d igit al board .
gg
is t he FPGA ver sion for t he digitizer 's chan n el 1- 2 an alog boar d.
rr
is t he FPGA ver sion for t he digitizer 's chan n el 3- 4 an alog boar d.
The FPGA ver sion s p rovide u sefu l in for m ation for in str u men t ser vicin g.
MAC Address
Then , t he in str umen t will disp lay its fixed MAC ad dr ess (a
typ ical examp le is shown ; ever y LAN in str u men t has a MAC add r ess
u n iqu ely id en tifyin g its LAN in ter face):
Idle State Message
Up on comp letin g t he p ower- on sequen ce, t he
in str umen t disp lays t his id le st at e message. If a LAN con n ection is
est ablished , the in str u men t also d isp lays th e IP add r ess:
In LAN con figu r ation s with ou t a DHCP ser ver, allow ap pr oximately 60
secon ds for th e Au tomatic IP featur e t o assign an IP ad dr ess to th e
d igitizer.
A br ief p ower on self t est occur s au tomatically du r in g power- on . If t his
t est fails, th e
an n u n ciat or on t he r igh t side of t he disp lay tu r n s on
when t he d igitizer en ter s t he Idle st ate. In t hat case, you may in itiat e a
mor e complete self test as d escr ibed in
Self Test Pr oced ur es
on
p age 76.
Should a LAN commu n ication er r or occur, t her e will be an ast er isk
ben eat h t he er r or an n u n ciator, on t he secon d display lin e, as follows:
Power-Down Sequence
To return the instrument to stand-by mode, press .