Calibration and Adjustment
4
Agilent L4532A/ 34A Service Guide
63
5
Select t he
Acquire Data
t a b.
a
Click t he
View Measurements...
bu tt on . Delete any in st alled
measu r emen ts.
b
Click
New
an d select Vr ms for each digitizer chan n el. Set th e Ar ea
t o Win dow an d choose AC cou p lin g. Click t he OK bu tt on .
c
Click th e
Edit..
button an d set t he St ar tin g Offset t o 0 an d Win d ow
Size t o 40000. Click th e OK bu tt on .
d
Leave t he In st alled Measu r emen ts win dow op en .
Procedure
1
On th e
Acquire Data
t ab, click t he
Run Once
button .
a
Not e t he Vr ms m easu r emen ts for all t he ch an n els in th e In st alled
Measu r emen ts win d ow.
b
Comp ar e t he measu r ed valu es t o t he on es shown in th e t able below.
2
Use t he
Configure Digitizer
t ab to chan ge th e Ran ge as in dicated in th e
t able below.
3
Rep eat st ep s 1 an d 2 for each r an ge in th e t able.
4
If any values ar e out of r an ge, r ep air may be r equ ir ed .
Input
Range
Error from 0
50 Termination
250 mV
0.00020 Vrms
500 mV
0.00024Vrms
1 V
0.00032Vrms
2 V
0.00050 Vrms
4 V
0.0036Vrms
8V
0.0064 Vrms
16 V
0.0128 Vrms