52
Agilent L4532A/ 34A Service Guide
4
Calibration and Adjustment
.
3
Make a con n ection t o t he digitizer u sin g t he web in t er face (see
p age 35
).
4
En ter th e Remot e Con trol Page an d select th e
Configure Digitizer
t ab. Set
th e followin g p ar amet er s an d click t he
Apply Configuration
bu tt on .
Chan n els:
Ran ge ±250 mV
Cou plin g DC
Filter LP_20_MHz
Acqu isition :
Samp le Rate 1 MSa/ s
Recor ds 1
Samp les/ r ecord 100000
Tr ig Hold off 0 sec
Pr e- tr ig Samp les/ r ecord 0
Tr ig delay 0
Tr igger :
Ar m sou rce Immed iate
Tr igger sou rce Immed iat e
L4534A 4-Channel