Calibration and Adjustment
4
Agilent L4532A/ 34A Service Guide
65
Calibration Count
You can r ead th e digitizer calibr ation cou n t t o deter min e how many
calibr ation s have been p er for med. Not e t hat your digitizer was calibr ated
befor e it left t he fact or y. When you r eceive you r d igitizer, be su r e t o r ead
t h e cou n t to det er min e its in itial valu e.
Th e calibr ation coun t in cr emen ts each time t he calibr ation con st an ts ar e
st or ed (
CAL:STOR
from the r emote in ter face). To r ead t he Calibr ation
Coun t u sin g t he Remote Con tr ol Page:
1
Select t he Utilities t ab.
2
Click t he Calibration... bu tton . Bot h t he calibr ation str in g an d
calibr ation cou n t can be r ead on th is scr een .
To r ead t he Calibr ation Cou n t u sin g th e Agilen t Con n ection Exp er t, u se
t he
CAL:COUN?
comman d (r efer to t he L4532A/34A Com m a n d
Refer en ce).
Calibration Security
Th is featu r e allows you to en ter a secu r ity cod e to p r even t acciden t al or
u n au th or ized adju stmen ts of t he digitizer. When you fir st r eceive your
d igitizer, it is secur ed . Befor e you can adju st t he digitizer, you mu st
u n secu r e it by en t er in g th e cor r ect secu r ity cod e.
Th e secu r ity code is set t o ATL4532 or ATL4534 wh en th e d igitizer is
ship p ed fr om th e factor y (d ep en din g on t he model). The secu r ity code
is stor ed in n on volatile memor y, an d does n ot chan ge when power
has been off, after a Fact or y Reset (
*RST
comman d ), or aft er an
digitizer Pr eset (
SYSTem:PRESet
comman d ).
The secu r ity cod e may con t ain up to u p t o 12 char act er s. You d o n ot
have t o u se all 12 char act er s bu t t he fir st char act er mu st always be a
lett er (A Z). The r emain in g 11 char acter s can be letter s, n u mber s
(0 9), or t he un der scor e char acter ("_"). Blan k sp aces ar e n ot allowed .
N O T E
The Calibration Count can be used to indicate that the digitizer calibration has
been tampered with. The calibration count is also incremented if the calibration
security code has been changed using the security bypass procedure (
To
Unsecure the digitizer Without the Security Code
on page 66).
N O T E
If you forget your security code, you can disable the security feature by following
the procedure below.