Calibration and Adjustment
4
Agilent L4532A/ 34A Service Guide
57
Input Value
Range
Filter
Quick
Check
Error from Input
-121.6 V
128 V
LP_20_MHz
0.3776 V
0 V
0.256 V
+121.6 V
0.3776 V
-121.6 V
LP_2_MHz
0.3776 V
0 V
0.256 V
+121.6 V
0.3776 V
-121.6 V
LP_200_kHz
Q
0.3776 V
0 V
Q
0.256 V
+121.6 V
Q
0.3776 V
-237.5 V
256 V
LP_20_MHz
0.5447 V
0 V
0.3072 V
+237.5 V
0.5447 V
-237.5 V
LP_2_MHz
0.5447 V
0 V
0.3072 V
+237.5 V
0.5447 V
-237.5 V
LP_200_kHz
Q
0.5447 V
0 V
Q
0.3072 V
+237.5 V
Q
0.5447 V