Calibration and Adjustment
4
Agilent L4532A/ 34A Service Guide
67
Programming the Adjustments
Comman ds u sed t o p er for m t he ad ju stmen ts ar e listed in
CALibration
su bsystem of th e L4532A/34A Com m a n d Refer en ce. A gen er al sequen ce
for u sin g t hese comman ds for a fu ll adju stmen t is as follows:
CAL:SECURE:STATE OFF,<secure_code> // unsecure for calibration
CAL:STATE ON
CAL:RANGE 0.25 // first range
CAL:RANGE:SOURCE UPP // sets high voltage output
// read the output voltage_hi
CAL:RANGE:SOURCE LOW // sets the low voltage output
// read the low voltage_lo
CAL:RANGE:CHAN? (@1:4), <voltage_hi>,<voltage_lo> // sets all channels
CAL:RANGE 0.5 // do the next range
CAL:RANGE:SOURCE UPP // sets high voltage output
// read the output voltage_hi
CAL:RANGE:SOURCE LOW // sets the low voltage output
// read the low voltage_lo
CAL:RANGE:CHAN? (@1:4), <voltage_hi>,<voltage_lo> // sets all channels
// repeat for ranges 1, 2, 4, 8, 16, 32, 64, 128, 256
// on the 32, 64, 128 and 256 ranges, the high and low input is 0
CAL:STORe // put the new calibration constants into memory
CAL:STATE OFF
CAL:SECURE:STATE ON,<secure_code>
Calibration Process
The followin g gen er al pr ocedu r e is t he r ecommen ded meth od to comp lete
a fu ll digitizer calibr ation u sin g t he Remote Access Pages.
1
Read
Test Con sid er ation s
on p age 48.
2
Per for m t he ver ification tests t o char acter ize th e d igitizer (in comin g
d at a).
3
On th e Remot e Access Page, select t he Utilities t ab. Click th e
Calibration... bu tton t o begin t he ad ju stmen ts. Note t he calibr ation
cou n t for th e digitizer 's main t en an ce r ecor d s.
4
Per for m t he ad ju stmen t p roced u r es (
Adju stmen t Pr oced ur es
on
p age 68).
5
Secu r e t he digitizer again st calibr ation .
6
Not e t he secu r ity code an d calibr ation cou n t in t he digitizer 's
main ten an ce r ecor ds.
N O T E
Note CAL:RANGE:CHAN? and CAL:STOR commands take several seconds to
execute. Be sure to set the remote time-out appropriately