Specifications
1
Agilent L4532A/ 34A Service Guide
7
G
eneral Specifications
Line Voltage:
Line Frequency:
Power Consumption:
1. CAT 1. IEC measurement category 1. Inputs may be connected only to circuits that are isolated from AC
mains.
Maximum Sample Rate
20 MSa/ s
Sample Resolution
16 Bits
Input Configuration
Isolated Inputs (each channel independently isolated)
Isolation voltage (low to chassis)
40 V
Maximum input (Hi to Low)
250 V
pk
1
Maximum Input Range
256 V
Input impedance
1 M | | 40 pF
Input coupling
DC or AC
AC cutoff frequency (-3 dB)
< 10 Hz
Input ranges
256 V, 128 V, 64 V, 32 V, 16 V, 8 V, 4 V,
2 V, 1 V, 500 mV, 250 mV
Overvoltage protection
Yes
Maximum Overvoltage Transient
400 V
pk
Analog bandwidth (-3dB)
20 MHz typical
Noise Filtering (2-pole Bessel)
200 kHz, 2 MHz typical
Power Requirements
100 to 240 VAC (universal)
50 Hz or 60 Hz
45 W (240 VA maximum)
EMC Conforms To:
IEC 61326-1:2005-12 (EU)
EN 61326-1:2006
ICES-001:2004 (Canada)
AS/ NZS CISPR 11:2004
Safety Conforms To
IEC/ EN 61010-1:2001( EU)
CAN/ CSA-C22.2 No. 61010-1-04 (Canada)
UL 61010-1 (2nd Edition) (US)
AS 61010.1:2003 (Australia/ New Zealand)