Specifications
1
Agilent L4532A/ 34A Service Guide
9
DC Accuracy
Total Specification ( % of r % of range )
1
1. 100,000 sample average @ 1 MSa/ s
2. For 1 V range and greater, typical offset with constant temperature is 0.01% of range.
3. Autozero done at room temperature within 24 hours.
23 C 5
C
T
autozero
3
C
2,3
Temp coefficient outside 18-28 C
Range
% of reading
% of range
% of range
% of reading
/ C
% of range/ C
250 mV
0.10
0.30
0.11
0.01
0.015
500 mV
0.10
0.20
0.06
0.01
0.010
4 V, 64 V
0.10
0.30
0.05
0.01
0.015
8 V, 128 V
0.10
0.20
0.04
0.01
0.010
1 V, 2 V, 16 V,
32 V, 256 V
0.10
0.12
0.04
0.01
0.010
Integral nonlinearity
5 LSB
Differential nonlinearity
1 LSB typical, no missing codes
Input bias current