Calibration and Adjustment
Input Channel Noise Functional Test (optional)
Description
This p roced ur e checks t he inp u t chan n el n oise of t he digitizer. This t est
does n ot measur e a war r an ted specification or char act er istic. However, it
is u sed to ver ify th at th e cir cu itr y in th e d igitizer is fu n ction in g p rop er ly.
The inp ut chan n el n oise fun ction al test con sists of d igitizin g a r ecor d of
d at a with t he inp ut chan n els ter min at ed with 50 t er min ation s an d
measu r in g th e Vr ms value of th e chan n el inpu t n oise. The inp u t n oise is
measu r ed by t he digitizer usin g t he On Boar d Measu r emen t cap ability t o
calculate t he Vr ms valu e of th e r ecor d. This measu r ed valu e is comp ar ed
t o t he exp ect ed valu e of the inp u t chan n el n oise fu n ction al test limit.
Equipment Setup
1
Read
Test Con sid er ation s
on p age 48.
2
Con n ect a 50 ter min ation t o each chan n el inp u t on th e d igitizer. You
may also u se a sin gle 50 t er min ation an d r ep eat t his test for each
chan n el. Do n ot con n ect t he chan n els in p ar allel t o a sin gle 50
ter min ation .
50 Terminations
3
Make a con n ection t o t he digitizer u sin g t he web in t er face (see
p age 35
).
4
En ter th e Remote Con trol Page an d select th e
Configure Digitizer
t ab. Set
th e followin g p ar amet er s an d click t he
Apply Configuration
bu tt on .
Chan n els:
Ran ge ±250 mV
Cou plin g DC
Filt er LP_20_MHz
Acqu isition :
Samp le Rate 20 MSa/ s
Record s 1
Samp les/ r ecord 40000
Tr ig Holdoff 0 sec
Pr e- tr ig Samp les/ r ecord 0
Tr ig d elay 0
Tr igger :
Ar m sou rce Immed iate