Calibration and Adjustment
4
Agilent L4532A/ 34A Service Guide
49
Performance Verification Tests
These p er for man ce ver ification tests ver ify the d igitizer is op er atin g
wit hin th e specification s sh own in
Chapt er 1
. Ther e ar e two ways th ese
p er for man ce t ests can be p er for med:
Usin g t he Web In t er face (as d escr ibed begin n in g on
p age 36
).
Cr eatin g your own p er for man ce t est p r ogr am usin g th e comman d s
descr ibed in th e Agilen t L4532A/34A Com m a n d Refer en ce (p r ovided
on t he CD wit h your d igitizer ).
The p r oced u r es in t his chap ter u se th e web in ter face. The same gen er al
p r ocedu r es can be adap ted t o u se eith er of t he ot her met hod s d escr ibed
above.
Self Test
A br ief p ower on self test occur s aut omatically wh en ever you cycle
p ower t o t he d igitizer. This limit ed test assu r es th at t he d igitizer is
cap able of oper ation . If t he power- on self test fails, t he
an n un ciat or
on t he r igh t sid e of t he disp lay tur n s on on ce t he d igitizer en t er s t he
Idle st ate. In t hat case, you may in itiat e a mor e comp let e self test as
d escr ibed below.
The comp lete self- test is an au t omated p rocedu r e th at systematically
exer cises t he p rocessor, digit al in fr astr uctu r e, memor y, an d an alog fron t
en d su bsyst ems. This p r oced ur e att emp ts t o isolate any failin g con dition s
fou n d to a p ar ticu lar assembly, t o facilit at e trou blesh ootin g an d r ep air.
The t est t akes ap p roximat ely 60 secon d s to comp lete. Su ccessfu l
execu tion of th e self- test p r ovid es a h igh degr ee of con fid en ce t hat th e
d igitizer is oper atin g n or mally.
Du r in g the t est t he d igitizer an alog in p uts ar e d iscon n ected in t er n ally
fr om th e r ear p an el BNC con n ect or s an d ar e con n ected to th e in t er n al
calibr ation sou r ce. Th is sou rce is t hen used to stimu lat e th e an alog inp ut
ch an n els wit h t est sign als. All of th e in ter n al sign al p ath s ar e tested for
p r op er op er ation . As t he d iffer en t sign al p at hs ar e con figu r ed for testin g,
t h e activation of th e r elays u sed t o swit ch th e sign al p at hs is au dible.
After t he t est p rocedu r e comp let es, t he an alog inp uts ar e r e- con n ected to
t he r ear pan el BNC con n ector s.
N O T E