6000 Series Oscilloscope User’s Guide
6
Displaying Data
Specifying the Channels Displayed After AutoScale
The
Channels
softkey selection determines which channels will
be displayed on subsequent AutoScales.
•
All Channels
— The next time you press
AutoScale
, all channels
that meet the requirements of AutoScale will be displayed.
•
Only Displayed Channels
— The next time you press
AutoScale
,
only the channels that are turned on will be examined for
signal activity. This is useful if you only want to view specific
active channels after pressing
AutoScale
.
Preserving the Acquisition Mode During AutoScale
The acquisition mode is normally switched to Normal when
AutoScale is performed. AutoScale can be set to leave the
acquisition mode unchanged if you prefer this option.
Choose Normal to make the oscilloscope switch to Normal
acquisition mode whenever the
AutoScale
key is pressed. This is
the default mode.
Choose Preserve to make the oscilloscope remain in the
acquisition mode you have chosen when the
AutoScale
key is
pressed.
Summary of Contents for InfiniiVision DSO6014L
Page 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Page 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Page 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Page 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Page 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Page 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Page 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Page 370: ......