156
6000 Series Oscilloscope User’s Guide
4
Triggering the Oscilloscope
In Asynchronous Mode
1
Press the
Baud
softkey to select a baud rate of
2.5 Mb/s
,
5 Mb/s
, or
10 Mb/s
.
2
Press the
Channel
softkey to select channel
A
or
B
.
In Synchronous Mode
1
Press the
FIBEX File
softkey to specify the location of the
FIBEX (Field Bus Exchange format) file that defines the
time-triggered timing schedule of the probed FlexRay bus.
2
Press the
Channel
softkey to select channel
A
or
B
.
In Asynchronous or Synchronous Mode
In addition to the logic
probe cable connection between the BUSDOCTOR and the
oscilloscope, you must make a LAN connection to allow the
oscilloscope to control the BUSDOCTOR. See also
N O T E
If the BUSDOCTOR 2 loses synchronization, then press the Mode key (in
the BUSDOCTOR 2 Menu) on the oscilloscope to cycle through the modes
and restore synchronization.
N O T E
Before powering-off the oscilloscope, change the BUSDOCTOR 2 mode to
PC.
If oscilloscope power is cycled when in Asynchronous or Synchronous
modes, communication with the BUSDOCTOR 2 will be lost. To restore
communication with the BUSDOCTOR 2 , cycle the BUSDOCTOR 2 power,
wait for its State indicator to turn from red to green, and then press the
Mode key (in the BUSDOCTOR 2 Menu) on the oscilloscope to cycle
through the modes.
Summary of Contents for InfiniiVision DSO6014L
Page 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Page 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Page 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Page 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Page 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Page 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Page 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Page 370: ......