148
6000 Series Oscilloscope User’s Guide
4
Triggering the Oscilloscope
5
Repeatedly press and release the
Baud
softkey to set the CAN
signal baud rate to match your CAN bus signal.
The CAN baud rate can be set to:
The default baud rate is 1 Mb/s
6
Repeatedly press and release the
Smpl Pt
softkey to select the
point between phase segments 1 and 2 where the state of the
bus is measured. This controls the point within the bit’s time
where the bit value is captured.
You can use the
Delayed
horizontal sweep mode for easier
navigation of the decoded data.
10 kb/s
50 kb/s
100 kb/s
500 kb/s
20 kb/s
62.5 kb/s
125 kb/s
800 kb/s
33.3 kb/s
83.3 kb/s
250 kb/s
1 Mb/s
One Bit
60%
70%
80%
Sample Point
N O T E
If the setup does not produce a stable trigger, the CAN signal may be slow
enough that the oscilloscope is AutoTriggering. Press the Mode/Coupling
key, then press the Mode softkey to set the trigger mode from Auto to
Normal.
Summary of Contents for InfiniiVision DSO6014L
Page 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Page 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Page 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Page 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Page 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Page 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Page 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Page 370: ......