196
6000 Series Oscilloscope User’s Guide
4
Triggering the Oscilloscope
To use USB triggering
USB trigger will trigger on a Start of Packet (SOP), End of
Packet (EOP) signal, Reset Complete (RC), Enter Suspend
(Suspend), or Exit Suspend (Exit Sus) on the differential USB
data lines (D+ and D-). USB Low Speed and Full Speed are
supported by this trigger.
1
Press the
More
key in the Trigger section of the front panel,
rotate the Entry knob until
USB
is displayed in the
Trigger
softkey, then press the
Settings
softkey to display USB trigger
menu.
2
Press the
Speed
softkey to select the speed of the transaction
being probed.
You may select Low Speed (1.5 Mb/s) or Full Speed (12 Mb/s).
3
Press the
D+
and
D–
softkeys to select the channel connected
to the USB signal D+ and D– lines. The
D+
and
D–
labels for
the source channels are automatically set.
D+,–
D–,+
Bus Idle
SOP
SE0 portion of EOP
End of packet trigger
Signal
speed
D+ source
channel
D– source
channel
Trigger on:
condition
Trigger level
or threshold
Currently selected
data source
USB trigger
Return to
previous menu
Summary of Contents for InfiniiVision DSO6014L
Page 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Page 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Page 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Page 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Page 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Page 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Page 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Page 370: ......