Triggering the Oscilloscope
4
6000 Series Oscilloscope User’s Guide
145
To use CAN triggering
Controller Area Network (CAN) trigger will trigger on CAN
version 2.0A and 2.0B signals. Setup consists of connecting the
oscilloscope to a CAN signal, using the
Settings
menu to specify
the event to trigger on, and using the
Signals
menu to specify the
signal source, baud rate, and sample point.
The oscilloscope will trigger on dominant-low CAN signals. If
you are connecting to the CAN signal using a differential probe,
connect the probe’s positive lead to the dominant-low CAN
signal (CAN_L) and connect the negative lead to the
dominant-high CAN signal (CAN_H).
The CAN trigger will trigger on the Start of Frame (SOF) bit of a
data frame. If the N5424A CAN/LIN Automotive Triggering and
Decode option is installed on your oscilloscope, the following
additional CAN trigger types will also be available: Remote
Frame ID (RTR), Data Frame ID (
~
RTR), Remote or Data Frame
ID, Data Frame ID and Data, Error Frame, All Errors,
Acknowledge Error, and Overload Frame.
A CAN message frame in CAN_L signal type is shown below:
1
Press the
More
key in the Trigger section of the front panel,
rotate the Entry knob until
CAN
is displayed in the
Trigger
N O T E
For CAN decode setup see
Arbitration
Field
Control
Field
CRC
Field
ACK
Field
Data
Field
Bus
Idle
Inter-
mission
EOF
SOF edge
Summary of Contents for InfiniiVision DSO6014L
Page 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Page 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Page 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Page 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Page 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Page 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Page 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Page 370: ......