Making Measurements
5
6000 Series Oscilloscope User’s Guide
247
Top
The Top of a waveform is the mode (most common value) of
the upper part of the waveform, or if the mode is not well
defined, the top is the same as Maximum. The Y cursor
shows the value being measured.
To isolate a pulse for Top measurement
The following figure
shows how to use delayed sweep to isolate a pulse for a
Top
measurement.
Figure 41
Isolating area for Top measurement