Triggering the Oscilloscope
4
6000 Series Oscilloscope User’s Guide
189
One example of triggering on a specific line of video is looking at
the vertical interval test signals (VITS), which are typically in
line 18. Another example is closed captioning, which is typically
in line 21.
1
Press the Trigger
More
key, then press the
TV
softkey.
2
Press the
Settings
softkey, then press the
Standard
softkey to
select the appropriate TV standard (NTSC).
3
Press the
Mode
softkey and select the TV field of the line you
want to trigger on. You can choose
Line:Field1
,
Line:Field2
, or
Line:Alternate
.
4
Press the
Line #
softkey and select the number of the line you
want to examine.
Figure 15
Example: Triggering on Line 136
Alternate Triggering
If Line:Alternate is selected, the oscilloscope will alternately trigger on the
selected line number in Field 1 and Field 2. This is a quick way to compare the
Field 1 VITS and Field 2 VITS or to check for the correct insertion of the half line at
the end of Field 1.
Summary of Contents for InfiniiVision DSO6014L
Page 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Page 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Page 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Page 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Page 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Page 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Page 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Page 370: ......